English | 中文
中文 » English » Browse By Category » Stimulators » Stimulators » Accupulser Signal Generator

Accupulser Signal Generator

/images/big693.jpg *The accuracy of digital electronics with the convenience of analog controls
Description Features
A pulse generator/stimulator combining the reproducibility and accuracy of digital electronics with the fine resolution and continuous adjustment possible with analog circuitry. All timing parameters are entered with ten-turn readable potentiometers and six-position range switches. Outputs are accurate to within 1% of the set value.

Pulses
Pulses can be created in continuous run, single-shot, or train/burst modes. Duration of the train/burst is easily controlled using the onboard envelope generator or by using either of two external gating inputs. Used in conjunction with the A360, A365, A385, or A395, bipolar pulses or trains may be easily produced. Output stimulus can be fed through Intra 767 or Duo 773 for iontophoresis. Footswitch allows hands-free operation.

Three separate outputs
Three separate outputs are available on the front panel. A Monitor output provides 10-15 V signals (up to 50 mA) for viewing the output on an oscilloscope or for controlling other devices. The stimulator’s signal, simultaneously available at the Isolator output, is sufficient to drive any WPI A300 Series stimulus isolator (A360, A365, or A385) and is also TTL and CMOS compatible. The Variable output can provide signals varying between ±10 V with a resolution of 1 mV. Separate variable outputs are provided for positive and negative signals.
Specifications
TIMING PARAMETERS: EVENT INTERVAL 100 ?s to1000 s*
TIMING PARAMETERS: EVENT DELAY 10 ?s to 100 s *
TIMING PARAMETERS: PULSE WIDTH 10 ?s to 100 s *
TIMING PARAMETERS: TRAIN DURATION (ENVELOPE) 100 ?s to 1000 s*
TIMING PARAMETERS: PULSE INTERVAL 20 ?s to 100 s*
OUTPUTS: SYNC 5 ?s, TTL, and 5 V CMOS compatible, 20 mA max.
OUTPUTS: MONITOR 10-15 V, 50 mA max.
OUTPUTS: ISOLATOR TTL & 5 V CMOS compatible, 20mA max.
VARIABLE (Pos or Neg):
VARIABLE (Pos or Neg): PULSED/DCLOW RANGE HIGH RANGE
VARIABLE (Pos or Neg): Range 0 to ±1 V 0 to ±10 V
VARIABLE (Pos or Neg): Resolution 1 mV 10 mV
NOISE:
NOISE: Pulsed at 100 kHz bandwidth <500 ?V
NOISE: DC Wide Band <500 ?V
OUTPUT IMPEDANCE <1 MΩ
INPUTS: EXTERNAL SYNC Accepts 1-?s minimum pulses TTL, CMOS compatible
INPUTS: EXTERNAL GATE Accepts 1-?s pulse to continuous TTL, CMOS compatible
POWER 95-130 V or 190-260 V, switch selectable single phase, 50/60 Hz
DIMENSIONS 17 x 5.25 x 10 in. (43 x 13 x 25 cm)
SHIPPING WEIGHT 14 lb (6.4 kg)
NOTES*Continuously variable in six ranges. All accuracies better than 1% of set value. 50kHz maximum pulse frequency.
Manuals and Resources Instruction Manual A310-IM-032807.pdf
References M.M. Lurtz, C.F. Louis. "Calmodulin and protein kinase C regulate gap junctional coupling in lens epithelial cells." Am J Cell Physiol 281. 2001: 972-981.

J.T. Porter, K.D. McCarthy. "Hippocampal Astrocytes In Situ Respond to Glutamate Released fromSynaptic Terminals." J of Neursciences 16. 1996: 5073-5081.

G. C. Churchill, M.M. Lurtz, C.F. Louis. "Ca2+ regulation of gap junctional coupling in lens epithelial cells." A, K/ {jusop; Ce;; {jusop; 287. 2001: 972-981.

M. Abo, J. Liang, L. Qian, J.D.Z.Chen. "Distension-Induced Myoelectrical Dysrhythmia and Effect of Intestinal Pacing in Dogs." Digestive Diseases and Sciences 45. 2000: 129-135.

A. Shafik, O. El-Sibai. "Rectal Pacing: Pacing Parameters Required for Rectal Evacuation of Normal and Constipated Subjects." J of Surgical Research 88. 2000: 181-185.

S. Karim, F. Gould, D.H. Dean. "Bacillus thuringiensis s-Endotoxin Proteins Show a Correlation in Toxicity and Short Circuit Current Inhibition Against Helicoverpa zea." Current Microbiiology 41. 2000: 214-219.

C-Y. Ho, Q. Gu, J-L. Hong, L-Y Lee. "Prostaglandin E2 Enhances Chemical and Mechanical Sensitivities of Pulmonary C Fibers in the Rat." Am J Respir Crit Care Med 162. 2000: 528-533.

X. Lin, J.Hayes, L.J.Peters, J.D.Z. Chen. "Entrainment of Segmental Small Intestinal Slow Waves with Electrical Stimulation in Dogs." Digestive Diseases and Sciences 45. 2000: 652-656.

X. Lin, L.J. Peters, J. Hayes, J.D.Z. Chen. "Entrainment of Intestinal Slow Wayves with Electrical Stimulation Using Intraluminal Electrodes." Annals of Biomedical Engineering 28. 2000: 582-587.

B.K. Shaw. "Involvement of a midbrain vocal nucleau in the production of both the acoustic and postural componenets of crowing behavior in Japanese quail." J Comp Physiology A 186. 2000: 747-757.

J.J. Ramirez, K.R.Bulsara, S.C. Moore, K. Ruch, W. Abrams. "Progressive Unilateral Damage of the Entorhinal Cortex Enhances Synaptic Efficacy of the Crossed Entorhinal Afferent to Dentate Granule Cells." J of Neuroscience 19. 1999: 1--6.
Ordering Information To place an order for this product, please contact the appropriate WPI office or distributor.

Customers can order direct from our China office:
WPI China
Address: Rm.20A, No.8 Dong Fang Rd, Liang Feng Building, Lu Jia Zui Financial District, Shanghai PRC
Phone:+86 21 68885517
Fax:+86 21 68885537
Code:200120
Accessories 3259- Footswitch for A310
2933- Rack Mount Kit, 5 1/4-in. high
Application Notes Application Notes No Application Notes available for this product at this time.
©Copyright 2008 World Precision Instruments, Inc. All Rights Reserved.