English | 中文
中文 » English » Browse By Category » Stimulators » Stimulus Isolators » Digital Linear Stimulus Isolator

Digital Linear Stimulus Isolator

/images/big694.jpg * Designed for use with the DS8000 digital stimulator
* High isolation and fast signal rise and fall times
* Two modes: current source and voltage source
* Designed for Muscle applications
Description Optical Coupling
DLS100 connects to the DS8000 via a flexible cable through which it receives power and stimulus signals in a digital format. Up to eight DLS100 isolators can be connected independently to one DS8000. Very high isolation is achieved through the use of optical coupling of the digital signal and a galvanically isolated DC power supply within the DLS100. Unlike some other multi-channel isolators, this digital isolator can be located at the site of the experiment, allowing the use of short connecting leads and thereby preserving high isolation and fast signal rise and fall times.

Two modes
The DLS100 operates in two modes: current source or voltage source. In the current source mode, the output current is proportional to the amplitude and polarity of the signal generated by the DS8000. In the voltage source mode, the output voltage is proportional to the amplitude and polarity of the signal generated by the DS8000. The DLS100 has user-selectable push-button switches that select different current or voltage ranges (the full-scale current from 100 nA to 10 mA, or the 10V and 100V full-scale, respectively). Three status indicators on the DLS100 indicate power on, activity (presence of signal from DS8000) as well as an alarm (over-range condition) indicator. Over-range can occur when the resistance of the load (the experiment) is too high for the current or voltage that is demanded from the DLS100.
Specifications
Curent Source Mode: Full-scale Current10 mA, 1 mA, 100 ?A, 10 ?A, 1 ?A, bipolar
Curent Source Mode: Compliance Voltage± 100 volts
Curent Source Mode: Output ImpedanceGreater than 100 Megohms
Curent Source Mode: Zero-signal LeakageLess than 0.01% of full-scale range setting ? 10 mV @ 100 V / 10,000 Ohms, 10 mA Scale = < 1 ?A Leakage
Curent Source Mode: LinearityBetter than 0.05% of full-scale range setting
Curent Source Mode: BandwidthRange and load dependant: 20 kHz with 10K load and 100 ?A or above range.
Voltage Source Mode: Full-scale Voltage± 100 volts
Voltage Source Mode: Maximum Current10 mA
Voltage Source Mode: Output ImpedanceLess than 1 ohm
Voltage Source Mode: Zero-signal OffsetLess than 1 mv
Voltage Source Mode: LinearityBetter than 0.05% of full-scale range setting
Voltage Source Mode: Bandwidth50 kHz
ISOLATION: ResistanceGreater than 1000 Megohms
ISOLATION: Capacitance Less than 10 pF, from output terminals to DS8000 and earth ground
POWER REQUIREMENTS +12 volts and +5 volts, supplied by DS8000
DIMENSIONS 14 x 9 x 3.5 cm (5.5 x 3.5 x 1.5 in.)
OUTPUT TERMINALS Mini-banana jacks
CONNECTING CABLE 150 cm (5 ft)
Manuals and Resources Instruction Manual DLS100-IM-100505.pdf
References Have you seen a reference to this product in a published paper? If so, email your contact information and the citation and we will send you a free gift!
Ordering Information To place an order for this product, please contact the appropriate WPI office or distributor.

Customers can order direct from our China office:
WPI China
Address: Rm.20A, No.8 Dong Fang Rd, Liang Feng Building, Lu Jia Zui Financial District, Shanghai PRC
Phone:+86 21 68885517
Fax:+86 21 68885537
Code:200120
Accessories 501670- Coaxial Adapter
83016- Replacement Cable, DLS100-to-DS8000
Application Notes Application Notes No Application Notes available for this product at this time.